Title | Microfluidic sterilization in a high temperature/pressure continuous flowing chip |
Authors | Zhang, Rui Wang, Baojun Chu, Ming Ding, Ran Wang, Yuedan Wang, Wei |
Affiliation | Institute of Microelectronics, Beijing, 100871, China School of Pharmaceutical Sciences, Beijing, 100871, China School of Basic Medical Sciences, Peking University, Beijing, 100871, China Pharmacy Departments, LiaoNingUniversity of Traditional Chinese Medicine, Liao Ning 110032, China National Key Laboratory of Science and Technology on Micro/ Nano Fabrication, Beijing, 100871, China |
Issue Date | 2013 |
Citation | 2013 17th International Conference on Solid-State Sensors, Actuators and Microsystems, TRANSDUCERS and EUROSENSORS 2013.Barcelona, Spain. |
Abstract | This paper reports the design, working principle, fabrication and the structure of a high temperature/pressure continuous flowing chip. The chip can reach a high pressure of 500kPa and temperature of 150??C, which can be used for portable cell culture system. The bacterial sterilization experiments are conducted with 108/ml E.coli and 107/ml Staphylococcus aureus flowing through the sterilization chip. The results show a sharp decrease of population proving the chip has an efficient and universal sterilization effects. ? 2013 IEEE. |
URI | http://hdl.handle.net/20.500.11897/410799 |
DOI | 10.1109/Transducers.2013.6627014 |
Indexed | EI |
Appears in Collections: | 基础医学院 |