Title Microfluidic sterilization in a high temperature/pressure continuous flowing chip
Authors Zhang, Rui
Wang, Baojun
Chu, Ming
Ding, Ran
Wang, Yuedan
Wang, Wei
Affiliation Institute of Microelectronics, Beijing, 100871, China
School of Pharmaceutical Sciences, Beijing, 100871, China
School of Basic Medical Sciences, Peking University, Beijing, 100871, China
Pharmacy Departments, LiaoNingUniversity of Traditional Chinese Medicine, Liao Ning 110032, China
National Key Laboratory of Science and Technology on Micro/ Nano Fabrication, Beijing, 100871, China
Issue Date 2013
Citation 2013 17th International Conference on Solid-State Sensors, Actuators and Microsystems, TRANSDUCERS and EUROSENSORS 2013.Barcelona, Spain.
Abstract This paper reports the design, working principle, fabrication and the structure of a high temperature/pressure continuous flowing chip. The chip can reach a high pressure of 500kPa and temperature of 150??C, which can be used for portable cell culture system. The bacterial sterilization experiments are conducted with 108/ml E.coli and 107/ml Staphylococcus aureus flowing through the sterilization chip. The results show a sharp decrease of population proving the chip has an efficient and universal sterilization effects. ? 2013 IEEE.
URI http://hdl.handle.net/20.500.11897/410799
DOI 10.1109/Transducers.2013.6627014
Indexed EI